Laboratory guide

How to read a peptide Certificate of Analysis

A peptide COA ties one sample to a stated method and result. Read it in the order the fields appear: whose sample, when, measured how, what was found, and how to check the document is genuine.

A phone scanning the QR code on a printed certificate

The short version: confirm the compound, lot and report number match your vial; check the dates; find the method behind every result; read purity, identity and content as three separate answers; read contaminant results against their reporting limits; look at the chromatogram; then verify the report with the laboratory that issued it, not with the seller.

This guide follows our example certificate, a marked specimen with illustrative figures, so you can open it in another tab and read along. The certificate of analysis glossary entry gives the short definition, and the annotated COA walkthrough works through the same document in a different format.

The layout of a peptide certificate

A Canada Peptide Testing certificate runs top to bottom in a fixed order. Each numbered block below links to its entry in the walkthrough.

  1. 1HeaderReport number CPT-EXAMPLE-001 and the date generated
  2. 2Sample identificationPeptide, client, lot number (or assigned reference); dates received, tested and reported
  3. 3HPLC-UV purity & content analysisExpected Content 10.00 mgMass 10.12 mgPurity 98.40%Fill Accuracy 101.2%
  4. 4Chromatogram, vial photograph and appearanceFig. 1 trace on released certificates; photo of the vial; appearance and cap colour
  5. 5Contaminant tables, when orderedEndotoxin or heavy metals: analyte, specification, result, unit, LOQ, status
  6. 6FooterResearch-use statement and the QR code for digital verification

Figures are the specimen’s illustrative values. No sample was tested to produce them.

Field-by-field walkthrough

For each field: what it means, what it shows on the example, what a good entry looks like, and the red flag to watch for.

Sample identity and client label

The sample identification table says which material the results belong to: the peptide named on the request, the client the report was issued to, and the lot number from the vial. When a vial carries no lot, the laboratory assigns a reference and the certificate labels it “Assigned Ref.” so it is never mistaken for a manufacturer’s lot. See sample ID.

On the example
Peptide BPC-157, client “Example client”, lot number EXAMPLE-BPC-001. Further down, a photograph of the vial and an appearance block (white lyophilized powder, dark grey cap) let you match the physical vial as well as the label.
Good entry
The compound and lot match the vial you are holding, character for character.
Red flag
No lot or sample identifier at all, or a lot that differs from the one on your vial. A certificate cannot be moved to another lot because the product name matches.

Report number and dates

The report number is the document’s address in the laboratory’s records, and it is what you look up to verify it. The dates record when the sample arrived, when it was analysed and when the result was reported. On a Canada Peptide Testing certificate they read Date Received, Date Tested and Date Reported, and the header shows the report number and the date the document was generated.

On the example
Report CPT-EXAMPLE-001: received Sep 3, tested Sep 8, reported Sep 9, 2026.
Good entry
Dates run in order (received, tested, reported) and fit the period when the lot was made and sold.
Red flag
No receipt date, a test date before the receipt date, or a report dated before the lot you are buying existed. An old certificate describes old stock.

Method lines

Every result needs the technique that produced it. For purity that means the chromatographic method and the detection wavelength, ideally with the column, gradient and run length. For content it means the reference standard and the basis; for endotoxin and metals, the assay or instrument technique. Without the method, a number cannot be compared with anything.

On the example
The results section is headed “HPLC-UV purity & content analysis”, and the Method & identity table below it lists RP-HPLC-UV at 214 nm on a C18 column (150 × 4.6 mm, 5 µm), water / acetonitrile with 0.1% TFA, a 5–60% acetonitrile gradient over 15 minutes, 1.0 mL/min, a 10 µL injection and a 20-minute run. Our methods page sets out the conditions behind each technique.
Good entry
A named method beside each result, with the conditions that change the number stated.
Red flag
“HPLC” with no wavelength, or a result with no method at all. These are the first two entries in the twelve red flags on a peptide COA.

Purity result

Area purity is the principal peak’s area as a percentage of all detected peak area at the stated wavelength. It describes the make-up of what the detector saw, not how much peptide the vial holds.

On the example
Purity 98.40% in the result strip, repeated as “HPLC area purity” in the summary above it.
Good entry
A percentage tied to a named method, a wavelength and the chromatogram it was calculated from.
Red flag
Purity presented as strength or dose, or a “99%+” with no trace and no wavelength. You cannot check a percentage whose evidence is withheld.

Identity result

Identity answers whether the sample is the compound the label names. It comes from comparing the sample with reference data for that compound, such as a reference spectrum. Canada Peptide Testing confirms identity against reference spectra using a selection of FTIR-ATR, Raman, capillary electrophoresis with UV detection and UV-Vis spectroscopy, chosen to suit the molecule.

On the example
Identity: consistent with reference, with the identity method given beside it as FTIR-ATR against a reference spectrum. The conclusion and the technique sit on the same line, so neither can be read without the other.
Good entry
The technique, the reference it was compared with, and a stated conclusion.
Red flag
“Identity: confirmed” with no method, or identity inferred from retention time alone. A matching retention time supports identity but does not settle it.

Content (mg per vial), when measured

Content is the measured mass of peptide in the vial, determined against a calibrated reference standard and expressed on a stated basis (peptide base or gross salt). On our certificates it fills the “Mass” cell, beside “Expected Content” (the label claim from your request) and “Fill Accuracy” (mass divided by label claim). Content is free on request when a suitable reference standard is available; when it was not measured, those cells read N/A.

On the example
Expected Content 10.00 mg, Mass 10.12 mg, Fill Accuracy 101.2%.
Good entry
A milligram figure with its basis, printed next to the label claim it is compared with.
Red flag
A content figure with no calculation basis, or a milligram figure that turns out to be the label claim multiplied by the purity. That is arithmetic, not a measurement.

Contaminant results and reporting limits

Endotoxin and heavy-metal results are separate tests. Each needs a unit (EU/mg or EU/mL for endotoxin), a reporting limit and, if it says pass or fail, a specification with its source. On our certificates they appear as tables with the columns Analyte, Specification, Result, Unit, LOQ and Status; an analyte below the limit reads “<” followed by the LOQ, with the status ND.

On the example
The example has no contaminant tables. They are added only when endotoxin or heavy-metals testing is ordered.
Good entry
A result such as “< 0.5 EU/mg”, with the limit and the specification named beside it.
Red flag
“Pass” with no specification, “not detected” with no reporting limit, or a result with no unit.

Chromatogram

The trace the purity figure was calculated from: detector response against retention time, with the principal peak marked. Our guide to reading a chromatogram covers it in depth.

On the example
The online specimen leaves the chromatogram out. On released certificates it sits beside the vial photograph as Fig. 1, with the detection wavelength in the caption; a representative trace is captioned as one, so it cannot pass for instrument data.
Good entry
Labelled axes, a marked principal peak, visible baseline noise, and a run that continues past the main peak.
Red flag
An unlabelled y-axis, a perfectly flat baseline, or a run that stops right after the principal peak, where late impurities would appear.

Sign-off and verification

This is how you know who stands behind the document and whether it has been altered. A Canada Peptide Testing certificate is released after staff review, signed, and printed with a QR code marked “Digital Verification” that opens its verification record. The record shows whether the document and the laboratory’s signatures check out, and says plainly if a certificate has been revoked. A correction is issued as a new version rather than an edit to the old one.

On the example
The specimen’s QR code leads back to the example page, and its status reads “Specimen” rather than “Verified”.
Good entry
A QR code or report number that opens a record on the laboratory’s own site, matching the document in every field.
Red flag
A QR code that leads to the seller’s site, a report number that resolves nowhere, or a record that differs from the PDF.

Purity, identity and content answer different questions

The most common misreading of a peptide COA is treating its one large percentage as the answer to every question. Three measurements sit on a full certificate, and each is blind to what the others measure.

QuestionResultWhat it cannot tell you
Is it the right compound?Identity, against reference data for the named compoundHow clean the material is, or how much is there
How clean is it?Area purity by HPLC, as a percentageWhich compound the main peak is, or how many milligrams the vial holds
How much is in the vial?Content in mg, against a reference standardWhether the material is free of impurities

Area purity is normalised. It is calculated from what the UV detector responded to, so salt, water and anything else that does not absorb at the chosen wavelength never enters the sum. That is why a vial can read 99% pure and hold a third of its labelled milligrams. Our guide to peptide purity versus content works through the difference with figures.

Identity is a comparison. A laboratory matches the sample against a reference for the compound the label names. A single sharp peak shows the sample is mostly one thing, not which thing; that is the job of the identity result. See purity versus identity for how the two methods cover each other’s blind spots.

Content is a measured amount. It needs a calibrated reference standard for the compound, which is why it is run on request rather than by default. Read it with its basis and with any measurement uncertainty in mind: a fill accuracy of 98% and one of 100% may not be a real difference.

Reporting limits: what “not detected” really says

Every method has a floor below which it cannot put a reliable number on a result. A reporting limit, usually set at or above the limit of quantitation, turns “not detected” into a statement you can evaluate: “< 0.5 EU/mg” means below 0.5, not zero. Compare the limit with the level you care about. A limit above your threshold means the test could not answer your question, whatever the status column says.

The same idea applies to purity. A related-substances list that stops at 0.1% of area and one that stops at 0.05% describe different levels of detail, so an empty impurity list is a statement about the threshold as much as about the sample.

Why one vial does not characterise a batch

A certificate describes the sample the laboratory received, on the date it was tested. Fill varies from vial to vial, and storage can change a sample after it leaves the manufacturer. A vendor’s certificate for one lot says nothing about a different lot, and even within one lot a single vial is one data point. Our note on what one vial tells you about a batch explains when testing several vials is worth it.

Red flags at a glance

Six of the patterns that should make you stop and ask questions. The full list, with what each one means, is in twelve red flags on a peptide COA.

Check a certificate step by step

If you have a certificate open now, the interactive COA checklist walks you through nine checks in order. The first four decide whether the document can be verified at all; the rest test whether the results mean what they appear to. Prefer paper? Download the printable COA checklist (PDF).

What a peptide COA does not establish

A certificate is a record of measurements on one sample. It does not establish clinical safety, efficacy, sterility, legal status or batch-wide uniformity, and a verification page confirms the document is genuine without widening what was tested. Canada Peptide Testing results are for research use only, and each certificate states these limits directly.

Need a certificate for a vial you have?

Request testing online. Staff review every request before you pay or ship anything, and the acceptance email carries the mailing address and packing steps. Content in mg per vial is free on request when a suitable reference standard is available. Already holding one of our certificates? Check it against the laboratory record.

Request testing Verify a certificate See the example certificate

Further reference

Sources and related guidance for this article:

01

What should a peptide certificate of analysis include?

A peptide COA should include a report number, the compound and lot it describes, the dates the sample was received, tested and reported, a named method beside every result, each result with its unit and basis, the chromatogram behind any purity figure, reporting limits for contaminant results, and a way to verify the document with the laboratory that issued it. A document missing several of these is a summary you cannot check.

02

What is a good purity on a peptide COA?

There is no single pass mark, because a purity figure only means something alongside the method that produced it. Suppliers commonly advertise 98% or higher by HPLC, but two figures are comparable only when the detection wavelength, the method and the chromatogram are shown. A high purity also says nothing about how many milligrams are in the vial.

03

Is purity the same as the amount of peptide in the vial?

No. Purity is the share of detected chromatographic signal in the main peak; content is the measured mass of peptide in the vial, in milligrams. A vial can read 99% pure and still hold far less than its label says. Only a content result compared with the label claim tells you whether the fill is right.

04

How can I tell if a peptide COA is real?

Check it with the laboratory that issued it, not with the seller. Scan the QR code or enter the report number on the laboratory’s own verification page, and confirm the compound, lot, dates and results match the document in your hand. If the report number does not resolve anywhere, or the record differs from the PDF, treat the certificate as unverified.

05

Does one COA cover every vial in a batch?

No. A certificate describes the one sample the laboratory received. Other vials from the same lot can differ in fill and, less often, in composition, so a batch-level conclusion needs several vials tested under a sampling plan.

06

What does “not detected” or “< LOQ” mean on a COA?

It means the result was below the method’s reporting limit, not that none was present. A result of “< 0.5 EU/mg” says the endotoxin level was under 0.5; the true value could be anywhere from zero up to that figure. A “not detected” with no limit printed beside it cannot be evaluated.

07

How do I get a COA for a peptide I bought in Canada?

Mail a sample to a Canadian laboratory that issues a verifiable certificate. At Canada Peptide Testing, staff review every request before you pay or ship anything. Peptide analysis, covering area purity by HPLC-UV and identity against reference spectra, is $250 CAD per sample, with volume rates from $220. Content in milligrams per vial is added free on request when a suitable reference standard is available. The released certificate carries a report number and QR code and is never listed publicly.